PRODUCTS

Catalog Number:FSA3055M
FS-Mouse PD-L1(Programmed Cell Death Protein 1 Ligand 1) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Mouse PD-L1(Programmed Cell Death Protein 1 Ligand 1) ELISA Kit Species Mouse
Uniprot ID Q9EP73 Alternative Names CD274, B7-H1, PD-L1, PDCD1L1, PDL1, B7 homolog 1
Detection method Sandwich Sensitivity 2.43 pg/mL
Standard 400pg/mL Detection Range 6.25-400pg/mL
Sample type Serum, Plasma and Tissue homogenate and Other biological samples;Sample volume:50μL
Reaction time 1.5H Research Area N/A
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse PD-L1 .Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse PD-L1 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse PD-L1 ,biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Mouse PD-L1 You can calculate the concentration of Mouse PD-L1 in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
400.00 2.570 2.512
200.00 1.825 1.767
100.00 1.314 1.256
50.00 0.685 0.627
25.00 0.313 0.255
12.50 0.230 0.172
6.25 0.108 0.050
0.00 0.058
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 13.034 66.571 167.000 16.634 62.286 165.000
Standard deviation 0.727 4.780 5.661 0.619 3.064 8.481
C V (%) 5.58 7.18 3.39 3.72 4.92 5.14
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 94-102 99
EDTA plasma (n=8) 96-103 99
Cell culture media (n=8) 89-98 93
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 93-101 96-104 95-104
Average (%) 98 99 99
1:4 Range (%) 97-106 91-97 97-105
Average (%) 102 94 101
1:8 Range (%) 89-99 95-104 97-106
Average (%) 94 99 101
1:16 Range (%) 94-101 96-104 96-103
Average (%) 97 101 99