PRODUCTS

Catalog Number:REA1026HF
MF-Human GROα/CXCL1(Growth Regulated Oncogene Alpha) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name MF-Human GROα/CXCL1(Growth Regulated Oncogene Alpha) ELISA Kit Species Human
Uniprot ID P09341 Alternative Names GROα/CXCL1
Detection method Sandwich Sensitivity 3.12 pg/mL
Standard 500pg/mL Detection Range 7.81-500pg/mL
Sample type Serum, Plasma and Tissue homogenate and Other biological samples;Sample volume:25μL
Reaction time 2.5H Research Area Cardiovascular, Signaling transduction, Cancer, Metabolism, Stem Cells
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to TGF-β1. Samples (or Standards) are added to the micro ELISA plate wells and combined with the specific antibody and biotinylated detection antibody specific for TGF-β1. Then Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain TGF-β1, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of TGF-β1. You can calculate the concentration of TGF-β1 in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
500.00 2.433 2.384
250.00 1.835 1.786
125.00 1.063 1.014
62.50 0.568 0.519
31.25 0.286 0.237
15.63 0.182 0.133
7.81 0.131 0.082
0.00 0.049
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 16.052 85.143 252.000 20.879 87.214 193.250
Standard deviation 0.743 3.244 13.986 1.558 3.131 10.532
C V (%) 4.63 3.81 5.55 7.46 3.59 5.45
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 93-100 96
EDTA plasma (n=8) 92-101 96
Cell culture media (n=8) 89-97 94
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 99-105 91-100 90-99
Average (%) 102 95 95
1:4 Range (%) 94-101 95-105 90-98
Average (%) 98 100 94
1:8 Range (%) 98-106 92-102 93-101
Average (%) 101 97 97
1:16 Range (%) 97-107 89-97 98-107
Average (%) 102 92 102