PRODUCTS

Catalog Number:RE10247
UA (Uric Acid) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name UA (Uric Acid) ELISA Kit Species Universal
Uniprot ID N/A Alternative Names UA
Detection method Competitive Sensitivity 331.25 ng/mL
Standard 40000ng/mL Detection Range 625-40000ng/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids, Sample volume: 50μL
Reaction time 1.5H Research Area Metabolic pathway; Signal Transduction; Cancer, Nutrition metabolism;Hormone metabolism; Signal transduction;Metabolic pathway; Enzyme & Kinase; Signal transduction;Metabolic pathway; Nutrition metabolism;Hormone metabolism;Cardiovascular,Signaling transduction,Neuroscience,Cancer,Metabolism; Enzyme & Kinase;Infection immunity;Endocrinology;Hormone metabolism; Metabolic pathway; Signal Transduction; Cancer
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with UA. During the reaction, UA in the sample or standard competes with a fixed amount of UA on the solid phase supporter for sites on the Biotinylated Detection Ab specific to UA. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of UA in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD
40000.00 0.259
20000.00 0.328
10000.00 0.566
5000.00 0.831
2500.00 1.255
1250.00 1.511
625.00 2.077
0.00 2.626
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 1725.517 6457.143 15440.000 1129.650 7142.850 22300.000
Standard deviation 75.405 401.634 1003.600 74.444 285.714 735.900
C V (%) 4.37 6.22 6.50 6.59 4.00 3.30
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 94-103 99
EDTA plasma (n=8) 93-100 97
Cell culture media (n=8) 94-102 97
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 91-100 95-103 92-99
Average (%) 96 98 96
1:4 Range (%) 93-102 97-104 97-106
Average (%) 97 100 101
1:8 Range (%) 96-104 96-105 98-104
Average (%) 101 100 101
1:16 Range (%) 89-97 96-106 89-98
Average (%) 94 101 94