PRODUCTS

Catalog Number:RE10234
Normetanephrine ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name Normetanephrine ELISA Kit Species Universal
Uniprot ID N/A Alternative Names Normetanephrine,NMN
Detection method Competitive Sensitivity 46.88 pg/mL
Standard 5000pg/mL Detection Range 78.13-5000pg/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids, Sample volume: 50μL
Reaction time 1.5H Research Area Metabolic pathway;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse PTX 3/TSG-14. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse PTX 3/TSG-14 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse PTX 3/TSG-14, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Mouse PTX 3/TSG-14. You can calculate the concentration of Mouse PTX 3/TSG-14 in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD
5000.00 0.278
2500.00 0.444
1250.00 0.780
625.00 1.165
312.50 1.538
156.25 1.881
78.13 2.122
0.00 2.459
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 85.130 706.090 1976.120 131.540 1306.780 1619.370
Standard deviation 4.350 46.730 100.840 4.370 62.760 90.390
C V (%) 6.55 6.94 5.26 7.49 7.89 5.62
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 87-99 93
EDTA plasma (n=8) 80-95 87
Cell culture media (n=8) 90-107 98
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 88-102 87-96 87-96
Average (%) 96 90 90
1:4 Range (%) 87-102 85-98 92-105
Average (%) 97 88 93
1:8 Range (%) 85-96 89-102 89-102
Average (%) 95 95 99
1:16 Range (%) 87-101 85-98 92-101
Average (%) 96 94 94