Catalog Number:RE10216
AICAR(AICA Riboside) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name AICAR(AICA Riboside) ELISA Kit Species Universal
Uniprot ID N/A Alternative Names 5-aminoimidazole-4-carboxamide-1-β-D-ribofuranoside
Detection method Sandwich Sensitivity 4.69 pg/mL
Standard 500pg/mL Detection Range 7.82-500pg/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids, Sample volume: 100μL
Reaction time 3.5H Research Area Metabolic pathway; Signal Transduction; Cancer, Nutrition metabolism;Hormone metabolism; Signal transduction;Metabolic pathway; Enzyme & Kinase; Signal transduction;Metabolic pathway; Nutrition metabolism;Hormone metabolism;Cardiovascular,Signaling transduction,Neuroscience,Cancer,Metabolism; Enzyme & Kinase;Infection immunity;Endocrinology;Hormone metabolism; Metabolic pathway; Signal Transduction; Cancer
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with AICAR. During the reaction, AICAR in the sample or standard competes with a fixed amount of AICAR on the solid phase supporter for sites on the Biotinylated Detection Ab specific to AICAR. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of AICAR in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
500.00 2.505 2.435
250.00 1.649 1.579
125.00 0.983 0.913
62.50 0.464 0.394
31.25 0.255 0.185
15.63 0.157 0.087
7.82 0.105 0.035
0.00 0.070
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 22.490 68.690 236.480 23.930 73.320 240.240
Standard deviation 1.120 3.430 11.820 1.200 3.670 12.010
C V (%) 5.43 2.68 2.83 5.76 2.81 2.99
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 85-97 91
EDTA plasma (n=8) 80-94 87
Cell culture media (n=8) 91-105 98
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 95-102 88-97 85-94
Average (%) 99 95 88
1:4 Range (%) 93-101 80-92 96-105
Average (%) 96 92 105
1:8 Range (%) 81-89 81-97 87-101
Average (%) 85 91 100
1:16 Range (%) 85-99 87-95 82-90
Average (%) 97 90 84