PRODUCTS

Catalog Number:RE3572H
Human SLC7A11( Solute Carrier Family 7, Member 11 )ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name Human SLC7A11( Solute Carrier Family 7, Member 11 )ELISA Kit Species Human
Uniprot ID Q9UPY5 Alternative Names CCBR1; xCT; Cystine/Glutamate Transporter,Cationic Amino Acid Transporter,y+ System; Amino acid transport system xc-; Calcium channel blocker resistance protein CCBR1
Detection method Sandwich Sensitivity 0.1 ng/mL
Standard 10ng/mL Detection Range 0.16-10ng/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL
Reaction time 3.5H Research Area Immunology
Test principle N/A
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
10.00 2.365 2.358
5.00 1.680 1.673
2.50 0.964 0.957
1.25 0.457 0.450
0.63 0.218 0.211
0.32 0.095 0.088
0.16 0.048 0.041
0.00 0.007
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 0.55 1.49 2.9 0.6 1.59 2.98
Standard deviation 0.03 0.07 0.15 0.03 0.08 0.15
C V (%) 6.98 4.6 2.58 7.29 4.71 2.6
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 83-95 89
EDTA plasma (n=8) 90-105 97
Cell culture media (n=8) 80-95 87
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 87-102 86-97 85-100
Average (%) 90 96 98
1:4 Range (%) 82-91 88-95 84-99
Average (%) 90 94 93
1:8 Range (%) 86-93 92-101 85-97
Average (%) 87 95 95
1:16 Range (%) 81-94 87-98 88-102
Average (%) 82 89 97