PRODUCTS

Catalog Number:RE10250
15-HETE(15-hydroxyeicosatetraenoic acid)ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name 15-HETE(15-hydroxyeicosatetraenoic acid)ELISA Kit Species Universal
Uniprot ID N/A Alternative Names 15 hydroxyeicosatetraenoic acid
Detection method Competitive Sensitivity 5.8 ng/mL
Standard 1000ng/mL Detection Range 15.63-1000ng/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids, Sample volume: 50μL
Reaction time 1.5H Research Area Signaling transduction;Cancer;Immunology;Metabolism
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with 15-HETE. During the reaction, 15-HETE in the sample or standard competes with a fixed amount of 15-HETE on the solid phase supporter for sites on the Biotinylated Detection Ab specific to 15-HETE. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of 15-HETE in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD
1000.00 0.163
500.00 0.397
250.00 0.657
125.00 1.155
62.50 1.629
31.25 1.822
15.63 1.962
0.00 2.451
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 26.483 181.714 435.50 41.103 126.000 492.00
Standard deviation 1.613 10.521 14.372 1.443 6.665 33.702
C V (%) 6.09 5.79 3.30 3.51 5.29 6.85
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 91-99 96
EDTA plasma (n=8) 86-95 91
Cell culture media (n=8) 94-103 99
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 90-97 96-104 89-95
Average (%) 94 99 92
1:4 Range (%) 92-98 93-102 92-100
Average (%) 95 97 96
1:8 Range (%) 91-98 95-104 91-98
Average (%) 94 100 94
1:16 Range (%) 91-100 98-105 97-105
Average (%) 96 101 101