PRODUCTS

Catalog Number:RE5442M
Mouse oxCL(oxidized cardiolipin) ELISA kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name Mouse oxCL(oxidized cardiolipin) ELISA kit Species Mouse
Uniprot ID N/A Alternative Names oxCL,oxidized cardiolipin
Detection method Sandwich Sensitivity 0.94 ng/mL
Standard 100ng/mL Detection Range 1.56-100ng/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL
Reaction time 3.5H Research Area Signal transduction;
Test principle N/A
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
100.00 2.363 2.311
50.00 1.561 1.509
25.00 0.958 0.906
12.50 0.526 0.474
6.25 0.271 0.219
3.13 0.156 0.104
1.57 0.095 0.043
0.00 0.052
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 3.22 19.71 33.69 3.52 19.99 34.01
Standard deviation 0.16 0.99 1.68 0.18 1 1.7
C V (%) 5.42 5.91 6.11 5.91 6.19 6.22
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 80-94 87
EDTA plasma (n=8) 89-101 95
Cell culture media (n=8) 82-95 88
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 80-92 82-94 88-95
Average (%) 84 89 88
1:4 Range (%) 80-93 82-94 98-107
Average (%) 91 90 102
1:8 Range (%) 91-98 95-102 90-99
Average (%) 95 100 98
1:16 Range (%) 89-102 87-103 88-97
Average (%) 90 89 89