Uniprot ID P12235 Alternative Names ADP/ATP translocase 1 ,ADP,ATP carrier protein 1,ADP,ATP carrier protein, heart/skeletal muscle isoform T1 ,ANT 1,Solute carrier family 25 member 4 ,Adenine nucleotide translocator 1 ,SLC25A4 ,ANT1 ,AAC1
Detection method Sandwich Sensitivity 0.09 ng/mL
Standard 10ng/mL Detection Range 0.156-10ng/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL
Reaction time 3.5H Research Area Metabolic pathway;Infection immunity;Hematology;Neuro science;
Test principle N/A
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
10.00 2.384 2.358
5.00 1.699 1.673
2.50 0.983 0.957
1.25 0.476 0.450
0.63 0.237 0.211
0.32 0.114 0.088
0.16 0.067 0.041
0.00 0.026
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 0.23 1.61 4.95 0.29 2.26 3.94
Standard deviation 0.01 0.12 0.24 0.01 0.07 0.24
C V (%) 3.21 5.97 2.92 6.88 3.13 4.44
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 80-95 87
EDTA plasma (n=8) 90-105 97
Cell culture media (n=8) 86-99 92
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 80-101 85-97 99-105
Average (%) 99 96 105
1:4 Range (%) 87-96 85-97 95-104
Average (%) 93 88 95
1:8 Range (%) 96-105 86-102 82-94
Average (%) 98 90 89
1:16 Range (%) 95-103 87-99 86-96
Average (%) 100 88 92