PRODUCTS

Catalog Number:RE1641B
Bovine TRH(Thyrotropin Releasing Hormone) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name Bovine TRH(Thyrotropin Releasing Hormone) ELISA Kit Species Bovine
Uniprot ID A6QM11 Alternative Names Thyrotropin-releasing hormone, TRH, Protirelin, TSH-releasing factor, Thyroliberin, Thyrotropin-releasing factor, TRF
Detection method Competitive Sensitivity 4.69 pg/mL
Standard 500pg/mL Detection Range 7.82-500pg/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Reaction time 1.5H Research Area CD & Adhesion molecule;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Bovine TRH. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to Bovine TRH. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Bovine TRH in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD
500.00 0.190
250.00 0.328
125.00 0.547
62.50 0.878
31.25 1.299
15.63 1.756
7.82 2.065
0.00 2.545
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 22.64 86.62 178.01 32.26 70.03 216.69
Standard deviation 0.31 3.67 10.91 1.33 3.49 8.96
C V (%) 7.01 5.69 2.69 5.28 7.08 3.69
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 85-99 92
EDTA plasma (n=8) 83-95 89
Cell culture media (n=8) 86-98 92
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 80-96 80-92 80-101
Average (%) 85 86 92
1:4 Range (%) 92-105 80-92 83-97
Average (%) 94 81 96
1:8 Range (%) 91-99 87-103 95-104
Average (%) 97 92 99
1:16 Range (%) 85-102 89-98 88-104
Average (%) 91 94 92