PRODUCTS

Catalog Number:RE1214M
Mouse GC(Glucagon) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name Mouse GC(Glucagon) ELISA Kit Species Mouse
Uniprot ID P55095 Alternative Names Carbohydrate-Rich Portion of Platelet Membrane Glycoprotein Ib Alpha Polypeptide
Detection method Competitive Sensitivity 18.75 pg/mL
Standard 2000pg/mL Detection Range 31.25-2000pg/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Reaction time 1.5H Research Area Metabolic pathway;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Mouse GC. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to Mouse GC. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse GC in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD
2000.00 0.209
1000.00 0.345
500.00 0.560
250.00 0.894
125.00 1.322
62.50 1.714
31.25 2.090
0.00 2.530
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 53.07 274.52 593.51 50.25 507.61 653.98
Standard deviation 1.00 18.08 26.71 1.33 23.17 27.63
C V (%) 5.19 5.40 5.03 7.37 6.69 6.34
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 80-95 87
EDTA plasma (n=8) 90-105 97
Cell culture media (n=8) 86-99 92
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 80-101 85-97 99-105
Average (%) 99 96 105
1:4 Range (%) 87-96 85-97 95-104
Average (%) 93 88 95
1:8 Range (%) 96-105 86-102 82-94
Average (%) 98 90 89
1:16 Range (%) 95-103 87-99 86-96
Average (%) 100 88 92
Assay Procedures