PRODUCTS

Catalog Number:FS1027H
FS-Human GCP-2(Granulocyte Chemotactic Protein 2) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Inquiry

Inquiry

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Human GCP-2(Granulocyte Chemotactic Protein 2) ELISA Kit Species Human
Uniprot ID P80162 Alternative Names chemokine (C-X-C motif) ligand 6 (granulocyte chemotactic protein 2); Chemokine alpha 3; CKA-3Granulocyte chemotactic protein 2; CXCL6; GCP-2; GCP2member 6 (granulocytechemotactic protein 2); GCP-2Small-inducible cytokine B6; Small inducible cytokine subfamily B (Cys-X-Cys), member b
Detection method Sandwich Sensitivity 0.50 pg/mL
Standard 250pg/mL Detection Range 3.91-250pg/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids;Sample Volume=50μL
Reaction time 1.5H Research Area Cancer, Immunology;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human GCP-2. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human GCP-2 are added to the micro ELISA plate wells. Human GCP-2 in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human GCP-2. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Human GCP-2. The concentration of Human GCP-2 in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
250.00 2.558 2.501
125.00 1.641 1.584
62.50 0.964 0.907
31.25 0.480 0.423
15.63 0.279 0.222
7.82 0.144 0.087
3.91 0.100 0.043
0.00 0.057
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 7.93 60.07 82.45 8.33 61.27 78.33
Standard deviation 0.62 3.45 5.05 0.50 4.82 5.39
C V (%) 7.82 5.74 6.12 6.00 7.87 6.88
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 87-95 91
EDTA plasma (n=8) 81-95 88
Cell culture media (n=8) 83-97 90
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 85-94 85-103 87-98
Average (%) 87 101 91
1:4 Range (%) 90-99 85-103 80-92
Average (%) 92 87 85
1:8 Range (%) 82-101 83-96 96-105
Average (%) 82 94 104
1:16 Range (%) 91-102 95-102 86-97
Average (%) 96 97 92
Assay Procedures