PRODUCTS

Catalog Number:RE3066Rb
Rabbit TGF-β2(Transforming Growth Factor β2) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

Delivery: Order now, ship in 3 days

Product Details
Product Name Rabbit TGF-β2(Transforming Growth Factor β2) ELISA Kit Species Rabbit
Uniprot ID N/A Alternative Names TGF-β2;TGFβ-2;TGF β2;TGF β 2;TGFβ 2;TGF-Beta2;TGFBeta-2;TGF Beta2;TGF Beta 2;TGFBeta 2;TGF-B2;TGFB-2;TGF B2;TGF B 2;TGFB 2;TGFB2, LDS4, TGF-beta2, transforming growth factor beta 2, G-TSF;TGF-B2; TGF-Beta2; G-TSF; LAP; Cetermin; Polyergin; Latency-associated peptide; BSC-1 cell growth inhibitor; Glioblastoma-derived T-cell suppressor factor
Detection method Sandwich Sensitivity 37.5 pg/mL
Standard 4000pg/mL Detection Range 62.5-4000pg/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids
Reaction time 3.5H Research Area N/A
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Rabbit TGF-β2 .Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Rabbit TGF-β2 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Rabbit TGF-β2 ,biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Rabbit TGF-β2 You can calculate the concentration of Rabbit TGF-β2 in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
4000.00 2.417 2.334
2000.00 1.577 1.494
1000.00 0.903 0.820
500.00 0.503 0.420
250.00 0.283 0.200
125.00 0.180 0.097
62.50 0.133 0.050
0.00 0.083 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 198.12 570.90 1347.79 210.71 579.43 1445.02
Standard deviation 9.91 28.55 67.39 10.54 28.97 72.25
C V (%) 5.74 2.84 7.00 5.85 2.84 7.26
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 80-95 86
EDTA plasma (n=8) 94-105 98
Cell culture media (n=8) 88-102 95
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 88-103 87-98 93-102
Average (%) 93 98 99
1:4 Range (%) 86-97 92-101 81-96
Average (%) 86 97 84
1:8 Range (%) 87-102 82-98 93-106
Average (%) 93 88 93
1:16 Range (%) 89-101 89-97 82-95