PRODUCTS

Catalog Number:RE10029
Biotin Quantitative Determination ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$459.00

$329.00

Delivery: Order now, ship in 3 days

Product Details
Product Name Biotin Quantitative Determination ELISA Kit Species Universal
Uniprot ID N/A Alternative Names Vitamin H, Coenzyme R, VB7, Vitamin B7, Biopeiderm
Detection method Competitive Sensitivity 5.35 pg/mL
Standard 500pg/mL Detection Range 7.81-500 pg/mL
Sample type Serum, Plasma and Other biological fluids; Sample Volume=50μL
Reaction time 0.75H Research Area Immunology
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Biotin. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Biotin are added to the micro ELISA plate wells. Biotin in samples (or standards) competes with a fixed amount of Biotin on the solid phase supporter for sites on the HRP linked detection antibody specific to Biotin. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of Biotin in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD
500.00 0.254
250.00 0.376
125.00 0.561
62.50 0.807
31.25 1.089
15.63 1.382
7.81 1.658
0.00 2.445
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 196.850 420.170 280.990 204.920 314.470 349.500
Standard deviation 20.980 46.360 145.090 26.110 57.700 180.600
C V (%) 4.350 4.500 4.240 3.860 5.940 5.430
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 90-105 97
EDTA plasma (n=8) 80-95 87
Cell culture media (n=8) 83-95 89
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 80-91 86-97 92-105
Average (%) 81 97 93
1:4 Range (%) 89-103 86-97 87-98
Average (%) 90 88 95
1:8 Range (%) 82-95 90-98 85-92
Average (%) 92 97 91
1:16 Range (%) 86-99 80-93 80-96
Average (%) 94 87 95