PRODUCTS

Catalog Number:FS3153H
FS-Human INS(Insulin) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$459.00

$329.00

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Human INS(Insulin) ELISA Kit Species Human
Uniprot ID P01308 Alternative Names IDDM2, ILPR, INS, Insulin, IRDN, MODY10, proinsulin
Detection method Sandwich Sensitivity 0.16 μIU/mL
Standard 20μIU/mL Detection Range 0.31-20μIU/mL
Sample type Serum, Plasma;Sample Volume=50μL
Reaction time 1.5H Research Area Enzyme & Kinase;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human INS. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human INS are added to the micro ELISA plate wells. Human INS in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human INS. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Human INS. The concentration of Human INS in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(μIU/mL) OD Corrected
20.00 2.852 2.851
10.00 1.645 1.644
5.00 0.879 0.878
2.50 0.449 0.448
1.25 0.228 0.227
0.63 0.103 0.102
0.31 0.052 0.051
0.00 0.001 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(μIU/mL) 0.534 6.507 6.446 15.460 18.211 19.783
Standard deviation 2.502 9.437 11.161 7.018 7.004 1.925
C V (%) 7.589 9.653 9.102 8.466 0.526 6.449
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 91-108 108
EDTA plasma (n=8) 93-98 94
Cell culture media (n=8) 90-93 90
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 88-103 85-116 103-115
Average (%) 103 111 114
1:4 Range (%) 87-116 97-108 86-115
Average (%) 107 103 115
1:8 Range (%) 95-104 100-117 93-115
Average (%) 101 102 107
1:16 Range (%) 95-112 94-110 93-104
Average (%) 98 107 99