PRODUCTS

Catalog Number:FS3130H
FS-Human CA242(Pancreatic Carcinoma Markers-CA242) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$459.00

$329.00

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Human CA242(Pancreatic Carcinoma Markers-CA242) ELISA Kit Species Human
Uniprot ID N/A Alternative Names CA242, Cancer Antigen 242, Carbohydrate Antigen 242
Detection method Sandwich Sensitivity 0.05 IU/mL
Standard 10IU/mL Detection Range 0.16-10IU/mL
Sample type Serum, Plasma;Sample Volume=50μL
Reaction time 1.5H Research Area Infection immunity;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human CA242. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human CA242 are added to the micro ELISA plate wells. Human CA242 in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human CA242. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Human CA242. The concentration of Human CA242 in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(IU/mL) OD Corrected
10.00 2.302 2.262
5.00 1.461 1.421
2.50 0.857 0.817
1.25 0.466 0.426
0.63 0.240 0.200
0.31 0.137 0.097
0.16 0.095 0.055
0.00 0.040 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(IU/mL) 0.950 1.804 2.965 9.781 8.484 3.149
Standard deviation 5.188 3.256 1.457 3.518 1.030 5.930
C V (%) 2.234 8.135 0.343 8.362 4.591 8.071
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 94-101 99
EDTA plasma (n=8) 89-107 104
Cell culture media (n=8) 81-112 103
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 85-95 92-104 100-108
Average (%) 87 93 102
1:4 Range (%) 90-111 103-113 84-111
Average (%) 104 105 111
1:8 Range (%) 97-107 94-118 95-112
Average (%) 97 111 102
1:16 Range (%) 85-108 98-104 89-110
Average (%) 86 101 103
Assay Procedures