Product Details
Product Name FS-Mouse IgG(Immunoglobulin G) ELISA Kit Species Mouse
Uniprot ID P01863 Alternative Names IgG, Immunoglobulin G, ImmunoglobulinG
Detection method Sandwich Sensitivity 0.09 ng/mL
Standard 100ng/mL Detection Range 1.56-100ng/mL
Sample type Serum, Plasma;Sample Volume=50μL
Reaction time 1.5H Research Area Neuroscience
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse IgG. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Mouse IgG are added to the micro ELISA plate wells. Mouse IgG in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Mouse IgG. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Mouse IgG. The concentration of Mouse IgG in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
100.00 2.594 2.593
50.00 2.035 2.034
25.00 1.156 1.155
12.50 0.768 0.767
6.25 0.448 0.447
3.13 0.251 0.250
1.56 0.136 0.135
0.00 0.001 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 88.516 31.065 76.557 38.552 46.422 85.866
Standard deviation 9.106 4.501 0.906 0.235 2.319 1.217
C V (%) 3.322 2.401 6.615 9.406 4.442 9.656
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 91-113 95
EDTA plasma (n=8) 97-101 97
Cell culture media (n=8) 84-96 92
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 86-104 99-111 103-110
Average (%) 90 102 109
1:4 Range (%) 98-117 86-109 102-114
Average (%) 112 102 111
1:8 Range (%) 91-110 93-110 90-109
Average (%) 104 105 90
1:16 Range (%) 99-116 91-107 101-106
Average (%) 105 94 105