PRODUCTS

Catalog Number:FS2840H
FS-Human NGAL(Neutrophil Gelatinase Associated Lipocalin) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$459.00

$329.00

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Human NGAL(Neutrophil Gelatinase Associated Lipocalin) ELISA Kit Species Human
Uniprot ID P80188 Alternative Names 24p3, 25 kDa alpha-2-microglobulin-related subunit of MMP-9, HNL, LCN2, lipocalin 2 (oncogene 24p3), lipocalin 2, Lipocalin2, Lipocalin-2, migration-stimulating factor inhibitor, MSFI, neutrophil gelatinase-associated lipocalin, NGAL, NGALlipocalin-2, Oncogene 24p3, p25, Siderocalin, Uterocalin
Detection method Sandwich Sensitivity 0.16 ng/mL
Standard 20ng/mL Detection Range 0.31-20ng/mL
Sample type Serum, Plasma;Sample Volume=50μL
Reaction time 1.5H Research Area Signaling transduction, Cell biology, Cancer, Immunology, Metabolism
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human NGAL ELISA Kit. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human NGAL ELISA Kit are added to the micro ELISA plate wells. Human NGAL ELISA Kit in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human NGAL ELISA Kit. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Human NGAL ELISA Kit. The concentration of Human NGAL ELISA Kit in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
20.00 3.202 3.200
10.00 1.971 1.969
5.00 1.064 1.062
2.50 0.592 0.590
1.25 0.325 0.323
0.63 0.169 0.167
0.31 0.083 0.081
0.00 0.002 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 11.365 1.546 11.266 16.641 7.269 3.262
Standard deviation 1.444 0.361 6.631 7.206 6.423 12.018
C V (%) 0.959 1.102 1.928 5.249 2.616 6.320
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 86-103 95
EDTA plasma (n=8) 97-108 105
Cell culture media (n=8) 91-110 110
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 92-106 89-108 94-118
Average (%) 103 107 110
1:4 Range (%) 89-108 83-111 89-116
Average (%) 102 101 114
1:8 Range (%) 90-107 99-108 103-118
Average (%) 96 106 117
1:16 Range (%) 92-105 94-115 87-107
Average (%) 100 100 104
Assay Procedures