PRODUCTS

Catalog Number:FS2831H
FS-Human ICAM-1/CD54(intercellular adhesion molecule 1) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$459.00

$329.00

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Human ICAM-1/CD54(intercellular adhesion molecule 1) ELISA Kit Species Human
Uniprot ID P05362 Alternative Names BB2, CD54 antigen, CD54, cell surface glycoprotein P3.58, human rhinovirus receptor, ICAM1, ICAM-1, intercellular adhesion molecule 1 (CD54), human rhinovirus receptor, intercellular adhesion molecule 1, Major group rhinovirus receptor, P3.58
Detection method Sandwich Sensitivity 18.75 pg/mL
Standard 2000pg/mL Detection Range 31.25-2000pg/mL
Sample type Serum, Plasma;Sample Volume=50μL
Reaction time 1.5H Research Area Signal transduction;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human ICAM-1/CD54. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human ICAM-1/CD54 are added to the micro ELISA plate wells. Human ICAM-1/CD54 in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human ICAM-1/CD54. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Human ICAM-1/CD54. The concentration of Human ICAM-1/CD54 in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
2000.00 3.576 3.575
1000.00 2.529 2.528
500.00 1.568 1.567
250.00 0.867 0.866
125.00 0.489 0.488
62.50 0.249 0.248
31.25 0.123 0.122
0.00 0.001 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 731.757 654.786 1324.367 424.267 1328.222 633.514
Standard deviation 6.910 5.434 4.874 13.499 12.223 12.003
C V (%) 5.175 8.788 9.598 8.928 9.430 6.598
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 89-96 93
EDTA plasma (n=8) 85-98 94
Cell culture media (n=8) 90-96 90
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 86-109 87-107 95-107
Average (%) 91 106 107
1:4 Range (%) 92-104 87-113 94-108
Average (%) 103 92 102
1:8 Range (%) 93-115 93-108 99-107
Average (%) 109 100 101
1:16 Range (%) 85-113 102-117 95-105
Average (%) 95 116 101
Assay Procedures