PRODUCTS

Catalog Number:FS2796M
FS-Mouse MMP-9(Matrix Metalloproteinase 9) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$459.00

$329.00

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Mouse MMP-9(Matrix Metalloproteinase 9) ELISA Kit Species Mouse
Uniprot ID P41245 Alternative Names 92 kDa gelatinase, 92 kDa type IV collagenase, CLG4B, EC 3.4.24, EC 3.4.24.35, Gelatinase B, GELB, macrophage gelatinase, MANDP2, matrix metallopeptidase 9, matrix metalloproteinase 9, matrix metalloproteinase-9, MMP9, MMP-9, type V collagenase
Detection method Sandwich Sensitivity 0.03 ng/mL
Standard 20ng/mL Detection Range 0.31-20ng/mL
Sample type Serum, Plasma;Sample Volume=50μL
Reaction time 1.5H Research Area Infection immunity;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse MMP-9. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Mouse MMP-9 are added to the micro ELISA plate wells. Mouse MMP-9 in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Mouse MMP-9. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Mouse MMP-9. The concentration of Mouse MMP-9 in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
20.00 2.947 2.945
10.00 1.973 1.971
5.00 1.188 1.186
2.50 0.707 0.705
1.25 0.384 0.382
0.63 0.202 0.200
0.31 0.100 0.098
0.00 0.002 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 11.894 1.828 18.398 19.789 3.532 6.651
Standard deviation 7.402 8.067 6.748 3.214 1.753 14.800
C V (%) 3.380 3.470 6.033 6.548 7.849 2.130
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 92-106 100
EDTA plasma (n=8) 84-106 102
Cell culture media (n=8) 91-95 94
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 89-102 96-109 86-110
Average (%) 89 98 87
1:4 Range (%) 98-105 99-108 90-104
Average (%) 99 103 96
1:8 Range (%) 100-105 89-118 85-118
Average (%) 103 116 98
1:16 Range (%) 86-114 94-110 96-111
Average (%) 96 105 108
Assay Procedures