PRODUCTS

Catalog Number:FS2746H
FS-Human MYO(Myoglobin) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$459.00

$329.00

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Human MYO(Myoglobin) ELISA Kit Species Human
Uniprot ID P02144 Alternative Names MB, MGC13548, Myoglobin, PVALB
Detection method Sandwich Sensitivity 0.94 ng/mL
Standard 100ng/mL Detection Range 1.56-100ng/mL
Sample type Serum, Plasma;Sample Volume=50μL
Reaction time 1.5H Research Area Immune molecule;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human MYO. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human MYO are added to the micro ELISA plate wells. Human MYO in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human MYO. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Human MYO. The concentration of Human MYO in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
100.00 3.328 3.327
50.00 2.567 2.566
25.00 1.515 1.514
12.50 0.759 0.758
6.25 0.327 0.326
3.13 0.138 0.137
1.56 0.063 0.062
0.00 0.001 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 93.340 35.308 9.087 34.173 15.578 32.362
Standard deviation 3.433 14.731 5.549 9.091 7.650 3.814
C V (%) 0.020 3.710 9.255 3.975 3.872 3.575
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 95-104 102
EDTA plasma (n=8) 97-103 101
Cell culture media (n=8) 87-111 106
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 93-98 91-117 96-118
Average (%) 98 112 111
1:4 Range (%) 90-107 91-105 95-117
Average (%) 104 91 108
1:8 Range (%) 84-105 98-114 83-118
Average (%) 94 103 110
1:16 Range (%) 101-104 103-108 87-115
Average (%) 101 104 107
Assay Procedures