PRODUCTS

Catalog Number:FS1739H
FS-Human SCF(Stem Cell Factor) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$459.00

$329.00

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-Human SCF(Stem Cell Factor) ELISA Kit Species Human
Uniprot ID P21583 Alternative Names c-kit Ligand, DCUA, DFNA69, DKFZp686F2250, familial progressive hyperpigmentation 2, FPH2, FPHH, KIT ligand, Kitl, KITLG, KL-1, Mast cell growth factor, MGF, MGFSHEP7, SCF, SCFStem cell factor, SFc-Kit ligand, SHEP7, SLF, steel factor
Detection method Sandwich Sensitivity 3.76 pg/mL
Standard 2000pg/mL Detection Range 31.25-2000pg/mL
Sample type Serum, Plasma;Sample Volume=50μL
Reaction time 1.5H Research Area Signal transduction;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human SCF. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human SCF are added to the micro ELISA plate wells. Human SCF in samples (or standards) combines with the coated antibody and HRP linked detection antibody special to Human SCF. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The OD value is proportional to the concentration of Human SCF. The concentration of Human SCF in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
2000.00 2.928 2.904
1000.00 1.739 1.715
500.00 0.939 0.915
250.00 0.523 0.499
125.00 0.278 0.254
62.50 0.156 0.132
31.25 0.091 0.067
0.00 0.024 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 1199.007 1875.491 369.613 893.014 259.799 647.123
Standard deviation 4.960 14.609 1.996 1.757 12.592 11.325
C V (%) 4.615 1.615 6.571 6.526 5.478 0.675
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 93-99 99
EDTA plasma (n=8) 96-107 98
Cell culture media (n=8) 84-103 86
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 88-95 98-108 93-110
Average (%) 90 101 108
1:4 Range (%) 98-107 86-115 85-109
Average (%) 99 101 91
1:8 Range (%) 95-114 88-116 91-115
Average (%) 99 96 93
1:16 Range (%) 96-114 99-105 83-112
Average (%) 101 99 104
Assay Procedures