Catalog Number:FS10177
FS-E2(Estradiol) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$459.00

$329.00

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-E2(Estradiol) ELISA Kit Species Universal
Uniprot ID N/A Alternative Names Oestradiol, E2, 17β-Estradiol, Estra-1,3,5(10)-triene-3,17β-diol
Detection method Competitive Sensitivity 11.25 pg/mL
Standard 800pg/mL Detection Range 12.5-800pg/mL
Sample type Serum, Plasma;Sample Volume=50μL
Reaction time 0.75H Research Area Signaling transduction;Cancer;Metabolism
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with E2. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for E2 are added to the micro ELISA plate wells. E2 in samples (or standards) competes with a fixed amount of E2 on the solid phase supporter for sites on the HRP linked detection antibody specific to E2. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of E2 in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD
800.00 0.134
400.00 0.296
200.00 0.579
100.00 1.062
50.00 1.617
25.00 1.973
12.50 2.117
0.00 2.375
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 790.107 264.434 433.132 232.548 677.945 495.630
Standard deviation 0.187 11.978 10.903 9.093 13.053 1.360
C V (%) 0.512 2.390 9.909 7.966 9.823 6.120
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 87-107 105
EDTA plasma (n=8) 88-107 107
Cell culture media (n=8) 90-105 92
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 91-97 83-118 84-114
Average (%) 94 102 85
1:4 Range (%) 97-118 85-115 85-117
Average (%) 116 96 116
1:8 Range (%) 91-113 91-108 87-106
Average (%) 94 93 91
1:16 Range (%) 100-111 101-104 83-107
Average (%) 109 101 88
Assay Procedures