PRODUCTS

Catalog Number:FS10133
FS-ALD(Aldosterone) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$459.00

$329.00

Delivery: Order now, ship in 3 days

Product Details
Product Name FS-ALD(Aldosterone) ELISA Kit Species Universal
Uniprot ID N/A Alternative Names ALD
Detection method Competitive Sensitivity 14.61 pg/mL
Standard 2000pg/mL Detection Range 31.25-2000pg/mL
Sample type Serum, Plasma;Sample Volume=50μL
Reaction time 0.75H Research Area Metabolic pathway;Endocrinology;Hormone metabolism;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with ALD. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for ALD are added to the micro ELISA plate wells. ALD in samples (or standards) competes with a fixed amount of ALD on the solid phase supporter for sites on the HRP linked detection antibody specific to ALD. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of ALD in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD
2000.00 0.133
1000.00 0.268
500.00 0.424
250.00 0.704
125.00 1.042
62.50 1.424
31.25 1.974
0.00 2.313
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 1804.979 1032.117 242.632 1999.256 1956.260 373.597
Standard deviation 11.273 8.612 8.334 9.153 12.900 14.576
C V (%) 3.308 5.725 6.028 3.827 8.856 9.447
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 85-104 87
EDTA plasma (n=8) 83-108 95
Cell culture media (n=8) 83-96 96
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 88-100 92-111 89-114
Average (%) 93 106 93
1:4 Range (%) 92-104 102-115 102-111
Average (%) 103 110 111
1:8 Range (%) 97-117 91-104 95-111
Average (%) 117 99 95
1:16 Range (%) 94-108 94-109 99-105
Average (%) 98 95 105
Assay Procedures