Product Details
Product Name MF-Human ET-1(Endothelin 1) ELISA Kit Species Human
Uniprot ID P05305 Alternative Names EDN1, ET1, HDLCQ7, PPET1, Preproendothelin-1
Detection method Sandwich Sensitivity 0.75 pg/mL
Standard 80pg/mL Detection Range 1.25-80pg/mL
Sample type Serum, Plasma and Tissue homogenate and Other biological samples;Sample volume:25μL
Reaction time 3H Research Area Signaling transduction, Cancer, Metabolism ;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human ET-1. Samples (or Standards) are added to the micro ELISA plate wells and combined with the specific antibody and biotinylated detection antibody specific for Human ET-1. Then Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human ET-1, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of Human ET-1. You can calculate the concentration of Human ET-1 in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
80.00 2.456 2.418
40.00 1.605 1.567
20.00 0.963 0.925
10.00 0.455 0.417
5.00 0.228 0.190
2.50 0.118 0.080
1.25 0.073 0.035
0.00 0.038 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 1.76 10.91 28.710 2.37 20.960 35.250
Standard deviation 0.07 0.53 1.67 0.10 0.89 1.32
C V (%) 4.57 4.90 3.22 5.99 3.39 4.81
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 91-105 98
EDTA plasma (n=8) 81-95 88
Cell culture media (n=8) 85-97 91
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 85-94 85-102 95-102
Average (%) 86 95 96
1:4 Range (%) 87-103 85-102 80-97
Average (%) 99 90 84
1:8 Range (%) 97-105 80-92 96-104
Average (%) 101 91 101
1:16 Range (%) 80-90 87-98 85-93
Average (%) 89 95 90