Catalog Number:RE2458RF
MF-Rat TNFRSF1B(Tumor Necrosis Factor Receptor Superfamily, Member 1B) ELISA Kit
Copy Product Information

Size:

48 T

96 T

Price:

$329.00

$459.00

Delivery: Order now, ship in 3 days

Product Details
Product Name MF-Rat TNFRSF1B(Tumor Necrosis Factor Receptor Superfamily, Member 1B) ELISA Kit Species Rat
Uniprot ID Q80WY6 Alternative Names TNFR2, CD12b, TBPII, TNF-R-II, TNF-R75, TNFBR, TNFR1B, TNFR8, p75, p75TNFR
Detection method Sandwich Sensitivity 9.38 pg/mL
Standard 1000pg/mL Detection Range 15.63-1000pg/mL
Sample type Serum, Plasma and Tissue homogenate and Other biological samples;Sample volume:25μL
Reaction time 3H Research Area Signaling transduction, Cell biology, Neuroscience, Cancer, Immunology;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Rat TNFRSF1B. Samples (or Standards) are added to the micro ELISA plate wells and combined with the specific antibody and biotinylated detection antibody specific for Rat TNFRSF1B. Then Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Rat TNFRSF1B, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of Rat TNFRSF1B. You can calculate the concentration of Rat TNFRSF1B in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
1000.00 2.461 2.430
500.00 1.581 1.550
250.00 0.888 0.857
125.00 0.453 0.422
62.50 0.244 0.213
31.25 0.138 0.107
15.63 0.087 0.056
0.00 0.031 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 25.70 204.64 493.100 29.28 203.410 448.740
Standard deviation 1.40 6.29 16.33 1.52 7.85 17.80
C V (%) 4.07 4.57 6.96 3.09 4.27 7.38
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 80-98 89
EDTA plasma (n=8) 82-94 88
Cell culture media (n=8) 84-98 91
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 80-92 87-102 89-97
Average (%) 81 90 93
1:4 Range (%) 88-96 87-102 95-104
Average (%) 88 95 100
1:8 Range (%) 80-91 96-105 87-96
Average (%) 85 103 95
1:16 Range (%) 85-93 85-93 88-101
Average (%) 88 91 96
Assay Procedures