Catalog Number:RE1464H
Human NTXⅠ(Cross Linked N-telopeptide of Type Ⅰ Collagen) ELISA Kit

Size:

48 T

96 T

Price:

$279.00

$399.00

Delivery: Order now, ship in 3 days

Product Details
Product Name Human NTXⅠ(Cross Linked N-telopeptide of Type Ⅰ Collagen) ELISA Kit Species Human
Uniprot ID N/A Alternative Names NTX-I, NTX1,NTX
Detection method Sandwich Sensitivity 1.88 ng/mL
Standard 200ng/mL Detection Range 3.13-200ng/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL
Reaction time 3.5H Research Area Signal transduction;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human NTXⅠ. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human NTXⅠ and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human NTXⅠ, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human NTXⅠ. You can calculate the concentration of Human NTXⅠ in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
200.00 2.347 2.284
100.00 1.553 1.490
50.00 0.922 0.859
25.00 0.439 0.376
12.50 0.260 0.197
6.25 0.160 0.097
3.13 0.113 0.050
0.00 0.063 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 4.76 40.21 83.32 3.39 29.01 60.78
Standard deviation 0.20 1.54 4.37 0.22 1.49 3.88
C V (%) 2.91 5.71 3.21 7.05 2.87 4.84
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 83-95 89
EDTA plasma (n=8) 90-105 97
Cell culture media (n=8) 80-95 87
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 93-102 87-98 91-99
Average (%) 93 94 98
1:4 Range (%) 81-96 92-104 83-89
Average (%) 92 104 84
1:8 Range (%) 93-106 82-96 88-104
Average (%) 103 94 104
1:16 Range (%) 82-95 95-102 97-103
Average (%) 83 97 102