PRODUCTS

Catalog Number:RE1076MF
MF-Mouse IL-12 p40(Interleukin 12 p40) ELISA Kit
Copy Product Information

Size:

48 T

96 T

Price:

$329.00

$459.00

Delivery: Order now, ship in 3 days

Product Details
Product Name MF-Mouse IL-12 p40(Interleukin 12 p40) ELISA Kit Species Mouse
Uniprot ID P43432 Alternative Names CLMF p40; CLMF; CLMF2; Cytotoxic lymphocyte maturation factor 40 kDa subunit; IL12 p40; IL-12 p40; IL-12 subunit p40; IL12B; IL-12B; IL-12BNK cell stimulatory factor chain 2; interleukin 12, p40; interleukin 12B (natural killer cell stimulatory factor 2, cytotoxic lymphocytematuration factor 2, p40); interleukin-12 beta chain; interleukin-12 subunit beta; natural killer cell stimulatory factor, 40 kD subunit; NKSF; NKSF2; NKSF2IL12, subunit p40
Detection method Sandwich Sensitivity 9.38 pg/mL
Standard 1000pg/mL Detection Range 15.63-1000pg/mL
Sample type Serum, Plasma and Tissue homogenate and Other biological samples;Sample volume:25μL
Reaction time 3H Research Area Cancer, Immunology;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse IL-12 p40. Samples (or Standards) are added to the micro ELISA plate wells and combined with the specific antibody and biotinylated detection antibody specific for Mouse IL-12 p40. Then Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse IL-12 p40, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of Mouse IL-12 p40. You can calculate the concentration of Mouse IL-12 p40 in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
500.00 2.301 2.234
250.00 1.386 1.319
125.00 0.766 0.699
62.50 0.431 0.364
31.25 0.269 0.202
15.63 0.167 0.100
7.81 0.114 0.047
0.00 0.067 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 70.25 89.43 157.66 155.55 66.02 157.84
Standard deviation 90.06 69.80 194.34 64.62 169.72 115.57
C V (%) 4.12 7.39 2.57 4.75 2.95 4.92
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 91-108 96
EDTA plasma (n=8) 88-115 94
Cell culture media (n=8) 89-110 93
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 85-104 82-96 85-101
Average (%) 89 94 97
1:4 Range (%) 80-92 85-96 84-98
Average (%) 90 90 94
1:8 Range (%) 83-95 87-101 85-93
Average (%) 92 101 86
1:16 Range (%) 86-97 89-97 86-98
Average (%) 88 95 87