PRODUCTS

Catalog Number:RE10005
F-TESTO(FreeTestosterone) ELISA Kit

Size:

48 T

96 T

Price:

$259.00

$359.00

Delivery: Order now, ship in 3 days

Product Details
Product Name F-TESTO(FreeTestosterone) ELISA Kit Species Universal
Uniprot ID N/A Alternative Names F-TESTO
Detection method Competitive Sensitivity 18.75 pg/mL
Standard 2000 pg/mL Detection Range 31.25-2000 pg/mL
Sample type Serum, Plasma, Tissue homogenates, Cell lysates, Cell culture supernates and Other biological fluids
Reaction time 1.5H Research Area Endocrinology;Reproductive science;Hormone metabolism;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with F-TESTO. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to F-TESTO. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of F-TESTO in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD
2000.00 0.204
1000.00 0.340
500.00 0.555
250.00 0.889
125.00 1.317
62.50 1.709
31.25 2.085
0.00 2.525
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 39.37 84.03 561.80 40.98 62.89 699.30
Standard deviation 4.20 9.27 29.02 5.22 11.54 36.12
C V (%) 5.96 4.41 3.09 5.50 5.74 4.09
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 82-94 88
EDTA plasma (n=8) 94-107 100
Cell culture media (n=8) 97-107 102
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 85-96 85-97 82-101
Average (%) 95 88 82
1:4 Range (%) 80-90 85-97 85-94
Average (%) 83 88 92
1:8 Range (%) 91-99 89-102 87-101
Average (%) 97 97 88
1:16 Range (%) 80-92 92-101 82-93
Average (%) 80 96 84
Assay Procedures