PRODUCTS

Catalog Number:RE10143
3-NT(3-Nitrotyrosine) ELISA Kit

Size:

48 T

96 T

Price:

$259.00

$359.00

Delivery: Order now, ship in 3 days

Product Details
Product Name 3-NT(3-Nitrotyrosine) ELISA Kit Species Universal
Uniprot ID N/A Alternative Names 3-Nitro-L-Tyrosine; 3-Nitrotyrosine; Nitrotyrosine
Detection method Competitive Sensitivity 1.35 ng/mL
Standard 300 ng/mL Detection Range 4.69-300 ng/mL
Sample type Serum, Plasma and Other biological fluids
Reaction time 1.5H Research Area Metabolic pathway;Infection immunity;Rheumatology;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with 3-NT. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to 3-NT. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of 3-NT in the samples is then determined by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD
300.00 0.235
150.00 0.357
75.00 0.535
37.50 0.819
18.75 1.128
9.38 1.496
4.69 1.769
0.00 2.517
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 5.91 12.61 84.27 6.15 9.43 104.90
Standard deviation 0.63 1.39 4.35 0.78 1.73 5.42
C V (%) 5.31 4.59 2.51 5.77 4.96 3.22
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 89-103 96
EDTA plasma (n=8) 83-97 90
Cell culture media (n=8) 86-99 92
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 86-93 92-101 80-96
Average (%) 87 94 92
1:4 Range (%) 93-101 92-101 95-103
Average (%) 98 100 101
1:8 Range (%) 85-98 87-98 80-96
Average (%) 91 92 95
1:16 Range (%) 90-101 85-92 93-106
Average (%) 93 87 99