Product Details
Product Name Human α-CTx(Alpha Crosslaps) ELISA Kit Species Human
Uniprot ID P50983 Alternative Names a-CTXI, aCTX-I, A-Cr, Type I Collagen C-Telopeptide-Related Fraction
Detection method Competitive Sensitivity 56.25 pg/mL
Standard 6000pg/mL Detection Range 93.75-6000pg/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Reaction time 1.5H Research Area Tumor immunity;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Human α-CTx. During the reaction, Human α-CTx in the sample or standard competes with a fixed amount of Human α-CTx on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Human α-CTx. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Human α-CTx in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD
6000.00 0.205
3000.00 0.383
1500.00 0.622
750.00 0.867
375.00 1.192
187.50 1.551
93.75 1.796
0.00 2.556
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 109.58 1038.44 1644.99 247.45 964.51 2450.25
Standard deviation 15.91 42.45 138.53 16.72 47.20 148.87
C V (%) 3.35 6.27 4.91 6.95 2.93 6.84
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 90-105 97
EDTA plasma (n=8) 88-102 95
Cell culture media (n=8) 82-97 89
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 89-98 87-98 82-96
Average (%) 93 90 88
1:4 Range (%) 81-94 87-98 81-94
Average (%) 90 96 89
1:8 Range (%) 96-105 92-105 93-107
Average (%) 99 102 99
1:16 Range (%) 92-103 95-107 93-102
Average (%) 102 102 99
Assay Procedures