PRODUCTS

Catalog Number:RE3270R
Rat LHRH(Luteinizing Hormone-Releasing Hormone) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$399.00

$279.00

Delivery: Order now, ship in 3 days

Product Details
Product Name Rat LHRH(Luteinizing Hormone-Releasing Hormone) ELISA Kit Species Rat
Uniprot ID N/A Alternative Names Gonadoliberin-1, Gonadoliberin I, Gonadotropin-releasing hormone I, GnRH-I, Luliberin I, Luteinizing hormone-releasing hormone I, LH-RH I
Detection method Competitive Sensitivity 1.88 mIU/mL
Standard 200mIU/mL Detection Range 3.13-200mIU/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Reaction time 1.5H Research Area Enzyme & Kinase;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Rat LHRH. During the reaction, Rat LHRH in the sample or standard competes with a fixed amount of Rat LHRH on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Rat LHRH. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Rat LHRH in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(mIU/mL) OD
200.00 0.197
100.00 0.333
50.00 0.548
25.00 0.882
12.50 1.310
6.25 1.702
3.13 2.078
0.00 2.518
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(mIU/mL) 6.81 28.24 65.93 12.17 33.30 108.80
Standard deviation 0.05 1.35 2.66 0.62 2.31 4.34
C V (%) 5.02 6.01 6.69 6.99 5.33 6.60
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 93-107 100
EDTA plasma (n=8) 87-99 93
Cell culture media (n=8) 92-107 99
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 92-101 86-101 92-105
Average (%) 98 92 101
1:4 Range (%) 82-98 86-101 86-93
Average (%) 89 89 91
1:8 Range (%) 89-97 86-101 91-101
Average (%) 91 89 92
1:16 Range (%) 85-93 87-96 87-98
Average (%) 89 94 97
Assay Procedures