PRODUCTS

Catalog Number:RE1213H
Human FLOT2(Flotillin 2) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$399.00

$279.00

Delivery: Order now, ship in 3 days

Product Details
Product Name Human FLOT2(Flotillin 2) ELISA Kit Species Human
Uniprot ID Q14254 Alternative Names ECS-1, ECS1, ESA, ESA1, M17S1
Detection method Sandwich Sensitivity 0.38 ng/mL
Standard 40ng/mL Detection Range 0.63-40ng/mL
Sample type Tissue homogenate;Sample Volume=100μL
Reaction time 3.5H Research Area Infection immunity;Urology;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human FLOT2. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human FLOT2 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human FLOT2, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human FLOT2. You can calculate the concentration of Human FLOT2 in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD Corrected
40.00 2.392 2.334
20.00 1.552 1.494
10.00 0.878 0.820
5.00 0.478 0.420
2.50 0.258 0.200
1.25 0.155 0.097
0.63 0.108 0.050
0.00 0.058 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 0.66 5.32 12.96 0.86 6.18 18.03
Standard deviation 0.04 0.47 0.69 0.06 0.46 0.89
C V (%) 5.85 3.94 2.66 5.64 3.17 5.47
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 80-92 86
EDTA plasma (n=8) 80-96 88
Cell culture media (n=8) 80-95 86
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 88-97 85-94 81-93
Average (%) 91 86 87
1:4 Range (%) 86-92 85-94 93-102
Average (%) 90 93 95
1:8 Range (%) 95-103 87-96 80-92
Average (%) 101 89 82
1:16 Range (%) 98-106 82-95 87-98
Average (%) 101 94 95
Assay Procedures