Catalog Number:RE3409M
Mouse Ang1-9(Angiotensin 1-9) ELISA Kit

Size:

48 T

96 T

Price:

$279.00

$399.00

Delivery: Order now, ship in 3 days

Product Details
Product Name Mouse Ang1-9(Angiotensin 1-9) ELISA Kit Species Mouse
Uniprot ID P11859(25~33aa) Alternative Names Angiotensin 1-9
Detection method Competitive Sensitivity 9.38 pg/mL
Standard 1000pg/mL Detection Range 15.63-1000pg/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Reaction time 1.5H Research Area Signal transduction;Neuro science;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Mouse Ang1-9. During the reaction, Mouse Ang1-9 in the sample or standard competes with a fixed amount of Mouse Ang1-9 on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Mouse Ang1-9. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Mouse Ang1-9 in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD
1000.00 0.280
500.00 0.380
250.00 0.574
125.00 0.835
62.50 1.130
31.25 1.525
15.63 1.839
0.00 2.503
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 24.57 148.72 301.02 35.48 177.00 342.90
Standard deviation 0.89 7.99 16.39 2.13 7.45 17.15
C V (%) 5.69 6.19 4.26 2.79 3.69 3.89
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 83-99 95
EDTA plasma (n=8) 85-97 91
Cell culture media (n=8) 97-105 101
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 80-97 90-103 92-101
Average (%) 94 95 98
1:4 Range (%) 86-92 90-103 95-106
Average (%) 90 91 101
1:8 Range (%) 81-93 97-106 85-104
Average (%) 81 103 91
1:16 Range (%) 89-97 82-90 90-98
Average (%) 92 89 91
Assay Procedures