PRODUCTS

Catalog Number:RE1645M
Mouse SCT(Secretin) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$399.00

$279.00

Delivery: Order now, ship in 3 days

Product Details
Product Name Mouse SCT(Secretin) ELISA Kit Species Mouse
Uniprot ID Q08535 Alternative Names SCT
Detection method Competitive Sensitivity 0.1 ng/mL
Standard 10ng/mL Detection Range 0.16-10ng/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Reaction time 1.5H Research Area cardiovascular disease
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Mouse SCT. During the reaction, Mouse SCT in the sample or standard competes with a fixed amount of Mouse SCT on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Mouse SCT. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Mouse SCT in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD
10.00 0.269
5.00 0.369
2.50 0.563
1.25 0.824
0.63 1.119
0.32 1.514
0.16 1.828
0.00 2.492
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 0.19 2.24 2.77 0.28 2.12 4.95
Standard deviation 0.03 0.08 0.15 0.03 0.11 0.19
C V (%) 6.21 4.70 6.54 6.42 4.34 3.73
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 81-95 88
EDTA plasma (n=8) 80-95 86
Cell culture media (n=8) 83-97 90
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 95-103 85-94 88-95
Average (%) 101 86 92
1:4 Range (%) 86-92 85-94 85-92
Average (%) 91 93 86
1:8 Range (%) 89-97 87-101 80-96
Average (%) 91 100 96
1:16 Range (%) 91-103 86-97 92-101
Average (%) 96 87 94
Assay Procedures