PRODUCTS

Catalog Number:RE1135M
Mouse ADH(Antidiuretic Hormone) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$359.00

$259.00

Delivery: Order now, ship in 3 days

Product Details
Product Name Mouse ADH(Antidiuretic Hormone) ELISA Kit Species Mouse
Uniprot ID P35455 Alternative Names AVP, ARVP, AVP-NPII, AVRP, VP, Vasopressin
Detection method Competitive Sensitivity 9.38 pg/mL
Standard 1000pg/mL Detection Range 15.63-1000pg/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Reaction time 1.5H Research Area Infection immunity;Immune molecule;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Mouse ADH. During the reaction, Mouse ADH in the sample or standard competes with a fixed amount of Mouse ADH on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Mouse ADH. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Mouse ADH in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD
1000.00 0.290
500.00 0.390
250.00 0.584
125.00 0.845
62.50 1.140
31.25 1.535
15.63 1.849
0.00 2.513
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 57.90 236.42 449.50 16.48 186.61 411.58
Standard deviation 1.77 7.71 21.38 0.05 9.39 19.30
C V (%) 5.05 5.75 2.56 2.87 3.16 2.93
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 84-106 100
EDTA plasma (n=8) 87-99 93
Cell culture media (n=8) 87-93 90
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 81-94 92-106 90-99
Average (%) 81 100 98
1:4 Range (%) 93-107 92-106 95-106
Average (%) 106 94 102
1:8 Range (%) 93-102 93-105 87-101
Average (%) 100 99 97
1:16 Range (%) 91-105 85-99 82-97
Average (%) 103 86 96
Assay Procedures