PRODUCTS

Catalog Number:RE2768H
Human CEA(Carcinoembryonic Antigen) ELISA Kit

Size:

48 T

96 T

Price:

$279.00

$399.00

Delivery: Order now, ship in 3 days

Product Details
Product Name Human CEA(Carcinoembryonic Antigen) ELISA Kit Species Human
Uniprot ID P06731 Alternative Names CD66E, CD66, CEACAM5, Carcinoembryonic Antigen-related Cell Adhesion Molecule 5, CEA, Carcinoembryonic Antigen
Detection method Sandwich Sensitivity 1.86 pg/mL
Standard 1000pg/mL Detection Range 15.63-1000pg/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL
Reaction time 3.5H Research Area Neuro science;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human CEA. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human CEA and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human CEA, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human CEA. You can calculate the concentration of Human CEA in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
1000.00 2.388 2.358
500.00 1.703 1.673
250.00 0.987 0.957
125.00 0.480 0.450
62.50 0.241 0.211
31.25 0.118 0.088
15.63 0.071 0.041
0.00 0.030 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 0.30 2.36 4.11 0.18 2.42 3.38
Standard deviation 0.02 0.10 0.17 0.02 0.11 0.17
C V (%) 2.92 3.45 4.35 4.55 6.54 6.50
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 80-95 86
EDTA plasma (n=8) 80-92 86
Cell culture media (n=8) 80-96 88
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 88-101 91-99 97-103
Average (%) 88 97 98
1:4 Range (%) 85-93 91-99 85-104
Average (%) 90 98 93
1:8 Range (%) 89-97 83-89 83-96
Average (%) 91 87 93
1:16 Range (%) 85-94 88-104 95-103
Average (%) 87 94 101
Assay Procedures