Product Name Human TNFSF14(Tumor Necrosis Factor Ligand Superfamily, Member 14) ELISA Kit Species Human
Uniprot ID O43557 Alternative Names CD258, HVEML, LIGHT, LTg, TR2
Detection method Sandwich Sensitivity 18.75 pg/mL
Standard 2000pg/mL Detection Range 31.25-2000pg/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL
Reaction time 3.5H Research Area Metabolic pathway;Cardiovascular biology;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human TNFSF14. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human TNFSF14 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human TNFSF14, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human TNFSF14. You can calculate the concentration of Human TNFSF14 in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
2000.00 2.311 2.284
1000.00 1.517 1.490
500.00 0.886 0.859
250.00 0.403 0.376
125.00 0.224 0.197
62.50 0.124 0.097
31.25 0.077 0.050
0.00 0.027 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 50.27 332.11 596.19 36.17 368.45 857.22
Standard deviation 1.95 18.27 40.24 2.27 17.94 46.03
C V (%) 6.31 6.93 3.80 6.71 7.41 6.41
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 95-105 100
EDTA plasma (n=8) 80-95 87
Cell culture media (n=8) 92-105 98
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 95-103 85-93 89-97
Average (%) 100 91 92
1:4 Range (%) 80-96 85-93 84-102
Average (%) 89 86 87
1:8 Range (%) 87-98 96-103 87-103
Average (%) 96 96 95
1:16 Range (%) 85-92 97-105 97-104
Average (%) 91 102 97