PRODUCTS

Catalog Number:RE2104H
Human L1CAM/CD171(L1-Cell Adhesion Molecule) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$399.00

$279.00

Delivery: Order now, ship in 3 days

Product Details
Product Name Human L1CAM/CD171(L1-Cell Adhesion Molecule) ELISA Kit Species Human
Uniprot ID P32004 Alternative Names L1CAM, CAML1, CD171, HSAS, HSAS1, MASA, MIC5, N-CAM-L1, N-CAML1, NCAM-L1, S1, SPG1, L1 cell adhesion molecule
Detection method Sandwich Sensitivity 56.25 pg/mL
Standard 6000pg/mL Detection Range 93.75-6000pg/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL
Reaction time 3.5H Research Area Enzyme & Kinase;Metabolic pathway;Hematology;
Test principle This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human L1CAM/CD171. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human L1CAM/CD171 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human L1CAM/CD171, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human L1CAM/CD171. You can calculate the concentration of Human L1CAM/CD171 in the samples by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(pg/mL) OD Corrected
6000.00 2.777 2.731
3000.00 1.715 1.669
1500.00 0.978 0.932
750.00 0.449 0.403
375.00 0.231 0.185
187.50 0.124 0.078
93.75 0.079 0.033
0.00 0.046 0.000
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(pg/mL) 160.14 1352.32 2377.15 153.14 1086.49 2179.79
Standard deviation 6.54 70.28 126.17 8.28 40.21 96.85
C V (%) 5.91 6.45 7.46 6.48 5.54 2.62
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 82-90 86
EDTA plasma (n=8) 84-92 87
Cell culture media (n=8) 95-104 98
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 88-96 85-94 97-105
Average (%) 92 86 98
1:4 Range (%) 90-101 80-92 84-96
Average (%) 94 82 88
1:8 Range (%) 93-102 87-101 87-98
Average (%) 102 91 94
1:16 Range (%) 85-94 82-90 95-102
Average (%) 94 85 100
Assay Procedures