Product Name Human TMSβ4(Thymosin Beta 4) ELISA Kit Species Human
Uniprot ID P62328 Alternative Names TMSB4X, FX, PTMB4, TB4X
Detection method Competitive Sensitivity 18.75 ng/mL
Standard 2000ng/mL Detection Range 31.25-2000ng/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Reaction time 1.5H Research Area Tumor immunity;Hepatology;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Human TMSβ4. During the reaction, Human TMSβ4 in the sample or standard competes with a fixed amount of Human TMSβ4 on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Human TMSβ4. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Human TMSβ4 in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD
2000.00 0.210
1000.00 0.346
500.00 0.561
250.00 0.895
125.00 1.323
62.50 1.715
31.25 2.091
0.00 2.531
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 118.08 347.43 743.86 26.44 324.94 570.98
Standard deviation 1.07 14.74 35.91 3.66 19.40 34.86
C V (%) 6.61 5.15 3.66 4.31 7.79 6.60
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 87-99 92
EDTA plasma (n=8) 88-107 94
Cell culture media (n=8) 97-105 101
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 90-103 95-106 82-98
Average (%) 92 101 88
1:4 Range (%) 97-106 95-106 80-97
Average (%) 99 96 91
1:8 Range (%) 82-90 85-104 86-92
Average (%) 86 93 88
1:16 Range (%) 92-101 90-98 81-93
Average (%) 94 91 82