PRODUCTS

Catalog Number:RE3102H
Human OB(Obestatin) ELISA Kit
Copy Product Information

Size:

96 T

48 T

Price:

$359.00

$259.00

Delivery: Order now, ship in 3 days

Product Details
Product Name Human OB(Obestatin) ELISA Kit Species Human
Uniprot ID Q9UBU3(76~98aa) Alternative Names OB
Detection method Competitive Sensitivity 0.47 ng/mL
Standard 50ng/mL Detection Range 0.79-50ng/mL
Sample type Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Reaction time 1.5H Research Area Metabolic pathway;Endocrinology;Hormone metabolism;
Test principle This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Human OB. During the reaction, Human OB in the sample or standard competes with a fixed amount of Human OB on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Human OB. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Human OB in tested samples can be calculated by comparing the OD of the samples to the standard curve.
Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

(ng/mL) OD
50.00 0.189
25.00 0.327
12.50 0.546
6.25 0.877
3.13 1.298
1.57 1.755
0.79 2.064
0.00 2.544
Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 2.65 10.87 13.40 0.26 13.06 22.67
Standard deviation 0.12 0.33 1.11 0.02 0.56 1.26
C V (%) 5.70 6.41 6.57 7.56 3.28 5.59
Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 85-99 92
EDTA plasma (n=8) 83-95 89
Cell culture media (n=8) 96-107 102
Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 86-94 82-95 96-102
Average (%) 92 93 96
1:4 Range (%) 91-105 82-95 83-97
Average (%) 102 91 88
1:8 Range (%) 87-99 83-96 84-99
Average (%) 91 91 91
1:16 Range (%) 95-100 88-95 95-102
Average (%) 96 95 101
Assay Procedures